4.1.2 User Calibration
User calibration is used when it is desired to calibrate the Indicator using a mass of known value from 10% to 100% of
the capacity of the scale base. To use this calibration feature, proceed as follows:
(FLASHING)
Press the Setup button, SCALE is displayed.
Press button UNTIL CAL is displayed.
Press Enter button, CAL TYPE is displayed.
Press Enter button, SPAN is displayed.
Press button to select user calibration. CAL
TYPE USEr is displayed.
Press Enter button; the display indicates the last
calibration mass value which was entered with the
first digit flashing. (Sample illustrates 10kg).
Press and or and enter the
desired mass value when using Model CD-31. The
numerical keypad can be used for Model CD-33. This
number must be at least 10% of the full span value.
Press Enter button, CLR PAN is displayed.
Press Enter button,WORKING is momentarily dis-
played, then PUT WT 10 kg is displayed.
Place specified calibration mass on pan.
Press Enter button, WORKING is displayed. After a
few seconds CAL SET is displayed, the display then
returns to WEIGH mode and indicates mass on pan.
• Remove calibration mass from the pan.
• User calibration is completed.