Page 1

I5S Front panel and PC board’

Page 2 - Instruction Manual

9Power OptionsPower OptionsPower OptionsPower OptionsPower OptionsThe I5S may be operated using the ACAdapter supplied, or 6 Alkaline “C” bat-teries (

Page 3

10kg Lights to indicate currentweighing unit is kilograms.g Lights to indicate currentweighing unit is grams.oz Lights to indicate currentweighing uni

Page 4 - TABLE OF CONTENTS

11Chapter 2Chapter 2Chapter 2Chapter 2Chapter 2Setup andSetup andSetup andSetup andSetup andCalibrationCalibrationCalibrationCalibrationCalibrationThe

Page 5

12Setup Mode and the Setup Mode and the Setup Mode and the Setup Mode and the Setup Mode and the IIIII5S Menu Structure5S Menu Structure5S Menu Struct

Page 6

13SCALEResetto FactorySettingsAveraging LevelHI or LOStability Level0.5, 1 or 3Unitslb ON or OFFkg ON or OFFoz ON or OFFg ON or OFFEndAuto Zero Tracki

Page 7 - Standard

14Setup SubmenuSetup SubmenuSetup SubmenuSetup SubmenuSetup Submenu1. With the indicator OFF, press andhold until SETUP is dis-played, then release i

Page 8 - Scale Base Connections

15ZeroZeroZeroZeroZeroZero specifies the percentage of fullcapacity load (2% or 100%) that may becleared by pressing .☞A 2% Zero setting is used for

Page 9

16Calibration UnitCalibration UnitCalibration UnitCalibration UnitCalibration UnitThe I5S may be calibrated using poundsor kilograms. Select the desir

Page 10 - Power Options

17Full Scale CapacityFull Scale CapacityFull Scale CapacityFull Scale CapacityFull Scale CapacityThe maximum capacity of the load re-ceiver is entered

Page 11 - Introduction and Installation

18Graduation SizeGraduation SizeGraduation SizeGraduation SizeGraduation SizeThis parameter specifies the incrementsin which the display will count.If

Page 12 - Setup and Calibration

1Instruction ManualINDICATOR StandardIndustrial ElectronicWeight IndicatorModel I5SOhaus CorporationOhaus CorporationOhaus CorporationOhaus Corporatio

Page 13 - 2 Setup and Calibration

19Calibration PointCalibration PointCalibration PointCalibration PointCalibration PointThe calibration point is the weight valuethat will be used to c

Page 14

20Menu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingThe menu lock

Page 15 - Setup Submenu

21Scale SubmenuScale SubmenuScale SubmenuScale SubmenuScale SubmenuIf you are already in the main menu, youcan repeatedly press until SCALEis display

Page 16

22Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory

Page 17

23To proceed to the next parameter (Sta-bility Level), press when AL isdisplayed.To exit the Scale submenu and storeyour settings, repeatedly press u

Page 18

24Stability LevelStability LevelStability LevelStability LevelStability LevelThe stability level specifies how much adisplayed weight may change (in g

Page 19

25Unit SelectionUnit SelectionUnit SelectionUnit SelectionUnit SelectionThe I5S permits you to display weight inpounds, kilograms, ounces and grams.To

Page 20

26Auto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero Tracking compensates forchanges in zero reading

Page 21

27Auto Off TimerAuto Off TimerAuto Off TimerAuto Off TimerAuto Off TimerThis parameter can be set to automati-cally turn the I5S off when it is idle f

Page 22 - Scale SubmenuScale Submenu

28Retain Zero DataRetain Zero DataRetain Zero DataRetain Zero DataRetain Zero DataSet this parameter to ON to save zerodata in permanent storage in ca

Page 23

2Unauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or

Page 24

29Calibration SubmenuCalibration SubmenuCalibration SubmenuCalibration SubmenuCalibration SubmenuThe calibration menu is used to performcalibration.I

Page 25

30RS232 SubmenuRS232 SubmenuRS232 SubmenuRS232 SubmenuRS232 SubmenuThe RS232 submenu only appears inthe main menu if the option is installed inthe ind

Page 26

31Legal For Trade SealingLegal For Trade SealingLegal For Trade SealingLegal For Trade SealingLegal For Trade SealingThe I5S has been tested and found

Page 27

3222222 Setup and CalibrationAudit Trail SealAudit Trail SealAudit Trail SealAudit Trail SealAudit Trail SealCalibration Event CounterCalibration Even

Page 28

33Setup and Calibration22222Paper SealPaper SealPaper SealPaper SealPaper SealApply the paper seal over one of the 6-32phillips pan head screws on the

Page 29

34Chapter 3Chapter 3Chapter 3Chapter 3Chapter 3OperationOperationOperationOperationOperationOnce the indicator has been setup and scale base connectio

Page 30 - Calibration Submenu

35Operation33333Selecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitRepeatedl

Page 31

3633333 OperationRe-ZeroingRe-ZeroingRe-ZeroingRe-ZeroingRe-ZeroingUse the RE-ZERO switch to reset thedisplay to zero. When weighing materialor objec

Page 32 - Legal For Trade Sealing

37Chapter 4Chapter 4Chapter 4Chapter 4Chapter 4TTTTTrrrrroubouboubouboubleshootingleshootingleshootingleshootingleshootingThe information in this chap

Page 33 - Audit Trail Seal

38TroubleshootingTroubleshootingTroubleshootingTroubleshootingTroubleshootingPROBABLE CAUSE(S)Adapter not plugged inor properly connected.Batteries de

Page 34

3TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSChapter 1 Introduction and Installation

Page 35 - 3 Operation

39Troubleshooting44444Error CodesError CodesError CodesError CodesError CodesDISPLAYEDMESSAGE DESCRIPTIONFlashes at 10 second intervals when batteries

Page 36 - Operation

4044444 Troubleshootingentered the Grad submenu, graduation selection will be limited to choices that donot exceed rated resolution.Err 23: Retain is

Page 37 - Re-Zeroing

41Chapter 5Specifications,ReplacementPartsand AccessoriesSpecifications, Replacement Parts and Accessories55555

Page 38 - Troubleshooting

4255555 Specifications, Replacement Parts and AccessoriesSpecificationsSpecificationsSpecificationsSpecificationsSpecificationsCapacity (Ib or kg) 5 t

Page 39

43Specifications, Replacement Parts and Accessories55555Replacement PartsReplacement PartsReplacement PartsReplacement PartsReplacement PartsOHAUSDESC

Page 40 - Error Codes

44Limited WarrantyLimited WarrantyLimited WarrantyLimited WarrantyLimited WarrantyOhaus products are warranted against defects in materials andworkman

Page 41 - 4 Troubleshooting

45Ohaus Corporation29 Hanover Road,Florham Park, NJ 07932, USATel: (973) 377-9000,Fax: (973) 593-0359With offices worldwide.P/N 77680-02 © Ohaus Corp

Page 42 - Replacement

4Chapter 2 Setup and Calibration (Cont.)RS232 Submenu ... 30Menu Protecti

Page 43 - SpecificationsSpecifications

5Chapter 1Chapter 1Chapter 1Chapter 1Chapter 1IntrIntrIntrIntrIntroduction andoduction andoduction andoduction andoduction andInstallationInstallation

Page 44

6IntroductionIntroductionIntroductionIntroductionIntroductionThis manual covers installation, operation and troubleshooting for the Ohaus I5SIndicato

Page 45 - Limited Warranty

7UnpackingUnpackingUnpackingUnpackingUnpackingUnpack the I5S indicator at the installation site and verify that the following itemswere included in th

Page 46

8Scale Base Wire ConnectionsBottom ViewPC Board+ EXCITATION+ SIGNAL- EXCITATION- SIGNALGROUND4. Connect the scale base wires to theterminal block loca