Output FormatsOutput Formats
Output FormatsOutput Formats
Output Formats
Data output can be initiated in one of three ways: (1) By pressing PRINT; (2) Using
the Auto Print feature; (3) Sending a print command (“P”) from a computer.
Data output by the I150 appears in the following formats.
Full Display Data Output:Full Display Data Output:
Full Display Data Output:Full Display Data Output:
Full Display Data Output:
When “diSP” is selected for “Print” in the rS-232 submenu, and “nu” is set to OFF, full
display data is output in the format shown in the diagram.
Weight Net
or or
Field Pol # pcs sp Unit sp Gross Stb CR LF
# of Characters 171211111
FULL DISPLAY DATA
lb, kg, oz,
cs or PC
Blank = Stable
? = Unstable
N = Net
G = Gross
Blank = positive
- = negative
Weight/# pieces is right justified and includes decimal
point. If polarity is negative, minus sign appears one
position to the left of the weight.
XX:XX:XX (time)
XX/XX/XX (date)
1234.56 lb N (display data)
NOTE: If time and date are set to ON in the
RS-232 submenu, output will include
time and date first.
Numeric Only Data Output:Numeric Only Data Output:
Numeric Only Data Output:Numeric Only Data Output:
Numeric Only Data Output:
When “diSP” is selected for “Print” in the rS-232 submenu, and “nu” is set to ON, only
numeric display data is output.
Weight
or
Field # pcs CR LF
# of Characters 8 1 1
NUMERIC ONLY DISPLAY DATA
NOTE: If time and date are set to ON in the
RS-232 submenu, output will include
time and date first.
Weight/# pieces is right justified and
includes decimal point and polarity.
Page view 67