  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 69
70
SYMPTOM
Unit will not turn on.
Cannot rezero indicator,
or will not re-zero when
turned on.
Center of Zero display in-
dicator erratic or does not
appear with no load on
load receiver.
Cannot display weight in
desired weighing unit.
Cannot display setup pa-
rameters or weight read-
ings for second load re-
ceiver.
Unit will not change weight
reading.
REMEDY
Check power cord con-
nections. Power indica-
tor light should be lit.
Remove load on load re-
ceiver to less than en-
tered zero %.
Change allowable zero %
in “ZERO” parameter of
setup submenu.
Remove disturbances or
reduct motion.
Increase averaging level
in “A.L.” parameter of
scale submenu.
Enable desired unit (see
section on SCALE SUB-
MENU).
Only KG and LB are avail-
able when LFT is set to
ON.
Select “duAL” for “PLAtF”
parameter of setup sub-
menu.
Check connections for
second load receiver.
Select proper load re-
ceiver.
PROBABLE CAUSE(S)
Power cord not plugged
in or properly connected.
Load on load receiver ex-
ceeds allowable zero %
entered in “ZERO” pa-
rameter of setup sub-
menu.
Retain set to ON and re-
called zero value is not
equal to current load.
Load receiver motion or
disturbances exceed cen-
ter of zero criteria.
Desired unit not set to ON
in “SELECt” parameter of
scale submenu.
“PLAtF” parameter of
setup submenu is set to
“SingLE”, or second plat-
form is not properly con-
nected.
I150 is set for dual load
receiver operation. Cor-
rect load receiver not se-
lected.
TROUBLESHOOTINGTROUBLESHOOTING
TROUBLESHOOTINGTROUBLESHOOTING
TROUBLESHOOTING