3. The display will show “buSY” while
the I150 reads the first calibration
point, and then show “C XXXX” indi-
cating the value of a load which must
be applied to the load receiver.
If the value is greater than 4 digits,
the display will show “C Pt” in
place of the value.
If 3-point linearization is being per-
formed (a linearization point was
selected in the “SEtUP” submenu),
and the value is greater than 4
digits, then “L Pt” will be displayed.
4. Apply the required load to the load
receiver, and then press ENTER.
5. The display will again show “buSY”
while the I150 reads the calibration
point.
- If span calibration is being per-
formed (no linearization point
was selected in the “SEtUP” sub-
menu), proceed to step 9.
- If 3-point linearization is being
performed (a linearization point
was selected in the “SEtUP”
submenu), continue with step 6.
6. The display will show “C XXXX” indi-
cating the value of the next load
which must be applied to the load
receiver. “C Pt” will be displayed if
the value is more than 4 digits.
OR
OR
OR