  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 38
39
7. Apply the required load to the load
receiver, and then press ENTER.
8. The display will show “buSY” while
the I150 reads the last calibration
point.
9. “CALib” will be displayed when cali-
bration is completed and the calibra-
tion submenu will be returned.
If two load receivers are being used,
you can press DOWN to switch to the
second load receiver, and repeat the
procedure.
NOTE: If calibration is aborted at any
time, previous calibration data will re-
main.
Calibration can be aborted by pressing
CLRR or OFFF.
EndEnd
EndEnd
End
You must select “End” to exit the calibra-
tion submenu.
1. Press ENTR when “End” is displayed.
2. “CAL” will be displayed and the main
menu will be returned.
NOTE: Calibration data is permanently
stored in memory when you use “End” to
exit from the Main menu, or when the OFF
key is pressed.